Yes, TOS7200 is equipped with an ANALOG OUT terminal that outputs a logarithmically compressed voltage from 0 to 4 V based on the resistance measurement Rx:
- 1 MΩ: 0.30 V
- 10 MΩ: 1.04 V
- 100 MΩ: 2.00 V
- 1000 MΩ: 3.00 V
- 10,000 MΩ or higher: 4.00 V
*Output impedance: 1 kΩ
For more details, read the user’s manual – chapter of Specifications – Interface and other functions.