Item | Common to all models | ||
---|---|---|---|
Synchronization function (clock synchronization) | Overview | SYNC icon is displayed on the display when synchronization is established with the internal clock after connecting with other PXZ series using the EXT SYNC connector. | |
Sequence synchronization | Synchronization of the program start and step start. | ||
Measurement synchronization | Synchronization of the measurement start | ||
Load synchronization | Synchronization of load ON/OFF | ||
Sequence function | Operation mode | CV, CC, CR and CP modes | |
Maximum number of programs | 30 | ||
Maximum number of steps | 10000 | ||
Step execution time | 1 ms to 3600000 s | ||
Loop count | 1 to 100000, or infinite | ||
Sine Function | Operation mode | CV/CC mode | |
Frequency setting range | 1 Hz to 1000 Hz | ||
Frequency precision setting | 1 Hz to 10 Hz | 0.2Hz | |
12 Hz to 100 Hz | 2Hz | ||
120 Hz to 1000 Hz | 20Hz | ||
CV | Maximum setting | Setting range up to 105 % of rated voltage | |
Maximum offset setting | |||
CC | Maximum setting | Setting range up to 105 % of rated current | |
Maximum offset setting | |||
Pulse function | Operation mode | CV/CC/CR mode | |
Frequency setting range | 1 Hz to 1000 Hz | ||
Frequency precision setting | 1 Hz to 10 Hz | 0.01Hz | |
12 Hz to 100 Hz | 0.1Hz | ||
120 Hz to 1000 Hz | 1Hz | ||
CV | High level | Setting range up to 105 % of rated voltage | |
Low level | |||
CC | High level | Setting range up to 105 % of rated current | |
Low level | |||
CR | High level | Setting range up to 105 % of rated conductance | |
Low level | |||
Duty cycle | 2.5 % to 97.5 % | ||
Over current protection (OCP) delay function | Setting range | 1 ms to 2000 ms | |
Resolution | 1 ms | ||
Multichannel (VMCB) function | Connection between the master unit and a PC | LAN, USB, RS232C | |
Connection with slave units | LAN | ||
Measurement trigger | Measurement start condition (trigger source) | Conditions for starting measurement can be selected (when inputting from display, when inputting commands by remote control, when inputting signals by external control, when operating in synchronization, and when load off) | |
Number of measurements | 1 to 65536 | ||
Measurement delay time | Setting range | 0 s to 100 s | |
Resolution | 0.1ms | ||
Measurement interval | Setting range | 0.1 ms to 3600 s | |
Resolution | 0.1ms | ||
Measurement time | Setting range | 0.1 ms to 1 s | |
Resolution | 0.1ms | ||
I-V characteristic function | Operation mode | CV/CC mode | |
Number of setup items | 3 to 100 items (interpolated between points with straight lines) | ||
Preset memory | Number of memory entries | 20 | |
Saved setting | Values in CV, CC, CP, and CR modes, and protection function values | ||
Setup Memory | Number of memory entries | 21 | |
Saved setting | Load on/off, Input voltage value/Input current value/Input power value/Conductance value, Input mode, Response, Slew Rate, Priority operation mode (Priority when load is ON), Value of the pulse function (Duty, Frequency, High, Low), Value of the sine function (Amplitude, Frequency, Offset), Number of I-V characteristics (Count), Over voltage protection (OVP), Under voltage protection (UVP, UVP Enable), Over current protection (OCP, Delay), Over power protection (OPP), Line overvoltage protection (Line OVP), Measurement trigger settings (Source, Count, Delay, Enable, Timer), Integration settings (Gate, Reset) | ||
key lock | Level 1 | Load on/off and preset memory recall are available | |
Level 2 | Load on/off are available | ||
Level 3 | Load off is available | ||
Number of parallel operated load modules | Up to 25 units | ||
Pre-charge function*1 | Maximum settable voltage | 105% of voltage rating | |
Voltage setting accuracy | ±(0.2 % of setting + 0.1 % of rating) | ||
Current setting accuracy*2 | ±(1.0 % of rating) |
*1 Release the interlock. *2 Fixed set value of 5 % of rated current.