0

BIM1000 Series

BIM1000_General specifications

ItemBIM1030BIM1100
EnvironmentInstallation locationIndoors, 2000 m or less
Spec guaranteed rangeTemperature / Humidity18 °C to 28 °C (-4 °F to 158 °F) / 20 %rh to 80 %rh (no condensation)
Operating rangeTemperature / Humidity0 °C to 40 °C (32 °F to +122 °F) / 20 %rh to 80 %rh (no condensation)
Storage rangeTemperature / Humidity-10 °C to 60 °C (-4 °F to 158 °F) / 90 %rh or less (no condensation)
Power supplyInput voltage range / Input frequency range85 Vac to 264 Vac (100 Vac to 240 Vac)/47 Hz to 63 Hz
Rated power30VA
Isolation voltage± 300 V max± 1000 V max
Insulation resistance30 MΩ or more (500 Vdc)(between AC LINE and chassis)
Withstanding voltageBetween the AC LINE and the chassis1500 Vac for 1 minute, 10 mA or less
Between all the mea- surement terminals and the chassis2000 Vdc for 1 minute, 1 mA or less
Between all the mea- surement terminals and SIGNAL I/O2000 Vdc for 1 minute, 1 mA or less
External dimensions / Weight214(8.43)W×80(3.15)H×300(11.81)D mm(inches) (Does not include protrusions)/ Approx. 3 kg (6.6 lbs)
Electromagnetic compatibility (EMC) *1 *2Complies with the requirements of the following directive and standards. EMC Directive 2014/30/EU EN 61326-1 (Class A *3), EN 55011 (Class A *3, Group 1 *4), EN 61000-3-2, EN 61000-3-3
Safety *1Complies with the requirements of the following directive and standards. Low Voltage Directive 2014/35/EU *2 EN 61010-1 (Class I *5, Pollution Degree 2 *6), EN 61010-2-030, EN 61010-031

*1. Does not apply to specially ordered or modified products. *2. Only on models that have CE/UKCA marking on the panel. *3. This is a Class A instrument. This product is intended for use in an industrial environment. This prod- uct may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts. *4. This is a Group 1 instrument. This product does not generate and/or use intentionally radio-frequen- cy energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose. *5. This is a Class I instrument. Be sure to ground this product’s protective conductor terminal.The safety of this product is guaranteed only when the product is properly grounded. *6. Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes that only non-conductive pollu- tion will occur except for an occasional temporary conductivity caused by condensation.


Related specifications : BIM1000

Please feel free to contact us with any questions or concerns.

Download a Product CatalogView List