PWR-01 series turns off the output soon after the output is turned on under the CC load condition. “AL/oCP” is shown on the panel. What should I do?

カテゴリ-
INDEX

“oCP” means an overcurrent protection (OCP), so the OCP function is activated.
Set the OCP value higher than the output current.
You can also limit the output current not to exceed approx. 95 % of the OCP trip point from the CONFIG setting (CF22).

Please feel free to contact us with any questions or concerns.