0

PCR-WEA/WEA2_Output Voltage Stability (Phase Voltage)

ModelSingle-phase outputSingle-phase/three-phase switchable model
 PCR1000WEAPCR2000WEAPCR3000WEA2PCR6000WEA2PCR12000WEA2PCR18000WEA2PCR24000WEA2PCR30000WEA2PCR36000WEA2
 PCR6000WEA2RPCR12000WEA2RPCR18000WEA2RPCR24000WEA2RPCR30000WEA2RPCR36000WEA2R
Line regulation *1Within ±0.1 %
Load regulation *2Within ±0.1 V/ ±0.2 V(1 Hz to 100 Hz)
Within ±0.3 V/ ±0.6 V(100.1 Hz to 500 Hz)
Within ±1 V/ ±2 V(500.1 Hz to 1 kHz)
Within ±0.2 V/ ±0.4 V(1 Hz to 100 Hz)
Within ±0.3 V/ ±0.6 V(100.1 Hz to 500 Hz)
Within ±1 V/ ±2 V(500.1 Hz to 1 kHz)
Output frequency variation *3When the output voltage correction function is enabled : Within ±0.3 %(1 Hz to 1 kHz), Within ±10 %(1001 Hz to 5 kHz) When the output voltage correction function is disabled : Within -3 dB(5 kHz)
Ripple noise *4≤ 0.25 Vrms
Ambient temperature variation *5±100 ppm/ °C (TYP)
Total harmonic distortion *60.3 % and less(1 Hz to 100 Hz), 0.5 % and less(100.1 Hz to 330 Hz), 1.5 %/kHz and less(330.1 Hz to 5 kHz)
Transient response *7Response FAST : 40 μs(TYP)
Response speed Tr/Tf *8Response FAST : 40 μs(TYP) Response MEDIUM : 100 μs(TYP) Response SLOW : 300 μs(TYP)

* 1 With respect to changes in the rated range of input voltage. 2 With respect to 0 % to 100 % changes in the rating of output current. When the output phase voltage is between 80 V and 160 V (L range) or 160 V and 320 V (H range) and the load power factor is 1, and the response is FAST. At the output terminal block, when the compensation function is not used. * 3 Voltage variation over 40 Hz to 5 kHz in AC mode with 55 Hz as the reference. When the output phase voltage is between 80 V and 160 V or 160 V and 320 V and the load power factor is 1, and the response is FAST, at the output terminal block. * 4 5 Hz to 1 MHz components in DC mode. * 5 With respect to changes in the operating temperature range. When the output phase voltage is 100 V or 200 V, with no load. * 6 When the output phase voltage is between 80 V and 160 V or 160 V and 320 V and the load power factor is 1, and the response is FAST, at the output terminal block. * 7 When the output voltage is 100 V or 200 V, the load power factor is 1, and the output current changes from 0 A to the rated value and from the rated value to 0 A. * 8 At 10 % to 90 % of the output voltage.


Related specifications : PCR-WEA/WEA2 Series

Please feel free to contact us with any questions or concerns.

Download a Product CatalogView List
Request a QuotationContact Us