TOS5400_Withstanding Voltage Test Section-Output function

AC output section (ACW only)

ItemTOS5401TOS5402
Output range0.05kV to 5.00kV
Output setting accuracy±(2% of setting + 20V) (at no load)
Output setting range0.000kV to 5.500kV
0.000kV to 1.500kV (When Time Unit is set to Cycle and Phase Shift is set to 0°)
Output resolution1V
Max. rated output *1500VA (5kV/100mA)
Max. rated voltage5kV
Max. rated current100mA (when the output voltage is 0.5kV or higher)
Transformer rating500VA
Output voltage waveform*2WaveformsSine
Phase shiftThe output voltage waveform can start with a phase shifted approx. 90° from the zero-crossing point.
Distortion3% or less. (when the output voltage is 0.5kV or higher and no load or a pure resistive load is connected)
Frequency50Hz/60Hz
Frequency setting accuracy±0.5% (excluding during voltage rise time)
Voltage regulation±10% or less (when changing from maximum rated load to no load)
Input voltage variation±0.3%(5kV at no load; power supply voltage: 90V to 250V)
Short-circuit current200 mA or more (output voltage 1.0 kV or higher)
Output methodPWM switching

*1 When tests are performed consecutively, output time limit and rest time may become necessary depending on the upper limit setting.
*2Waveform distortions may occur if an EUT whose capacitance is dependent on voltage (for example, an EUT that consists of ceramic capacitors) is connected as the load. However, if the test voltage is 1.5kV, the effect of a capacitance of 1000pF or less can be ignored. Because the high-voltage power supply of the TOS54 series uses the PWM switching method, if the test voltage is 500V or less, the switching and spike noise proportions are large. The lower the test voltage, the greater the waveform is distorted.

DC output section (DCW only)

ItemTOS5401TOS5403
Output range0.05kV to 6.00kV
Output setting accuracy±(2% of setting + 20V) (at no load)
Output setting range0.000kV to 6.200kV
Output resolution1V
Max. rated output *150W (5kV/10mA)
Max. rated voltage6kV
Max. rated current10mA
Ripple (TYP)5kV no load50Vp-p
Max. rated load100Vp-p
Voltage regulation3% or less (when changing from maximum rated load to no load)
Short-circuit current (TYP)40mA (when generating 6 kV output)
Discharge functionForced discharge after test completion (discharge resistance: 125kΩ)*2 Maximum capacitance: 1μF*3
Discharge time: 0.0 to 100.0 seconds
Discharge function during interlock activation: ON/OFF

*1 When tests are performed consecutively, output time limit and rest time may become necessary depending on the upper limit setting.
*2 In the following conditions, charging current may activate a FAIL judgment or trigger the protection function. EUT capacitance: 1μF; Rise Time: 1.0 second or less; Judge Delay: 1.2 seconds or less.
*3 Maximum capacitance that can be discharged within 2 seconds in a DC withstanding voltage test. Tests can still be conducted beyond the maximum capacitance, but the discharge time will increase.

Common specifications

ItemTOS5401TOS5402TOS5403
Start voltage functionThe initial voltage at test start can be set to 1% to 99% of the test voltage. Resolution: 1%
Disabled if the Time Unit is set to Cycle.n/a
Limit voltage functionThe upper limit of the test voltage can be set.
ACW: 0.000kV to 5.500kV, DCW: 0.000kV to 6.200kV
ACW: 0.000kV to 1.500kV (When Time Unit is set to Cycle and Phase Shift is set to 0°)
Output voltage monitor functionIf the output voltage deviates more than ±(350V) form the set value, the output is turned off, and the protection function is activated.