TOS5400_Auto Test
This table is part of the specifications for:TOS5403
| Item | Specifications |
|---|
| Number of programs | 61 |
|---|
| Number of steps | 2 |
|---|
| Test mode | The same settings as for single test settings are possible, except for the following. The same test can be selected consecutively.- Limit Voltage is fixed to the maximum rated value per test mode.
- Test Time is fixed to ON.
- In IR test, Lower is fixed to ON (only for models that support IR testing).
|
|---|
| Fail judgment operation | The following operations are selectable:- Continue: All steps are executed even for a FAIL judgment.
- Abort: Automatic test ends for a FAIL judgment.
|
|---|
| Interval time between steps | 0.1s to 100s |
|---|
| Step start condition | The following operations are selectable:- Immediate: When the INIT or INIT:TEST command is received via remote control, starts all steps consecutively.
- BUS: Step 1 can be started using the software trigger from remote control. After it is completed, the operation pauses. When the command is then received again, Step 2 starts.
- External: Step 1 starts upon performing the start operation. After a step is completed, the operation is paused. When you perform another start operation, step 2 starts.
- Once: When you perform a start operation, all steps are executed consecutively.
|
|---|