Operating voltage | 1.5 Vdc to 150 Vdc *1 |
---|---|
Current | 400 A |
Power | 2000 W |
Current setting accuracy *2 | ±(1.2 % of set + 1.1 % of f.s) *3 |
*1 Minimum voltage at which the current starts flowing to the PLZ2004WB is approx. 0.3 V. The PLZ2004WB detects no signal at an input voltage less than or equal to approx. 0.3 V and an input current less than or equal to approx. 1 % of the range rating. If the input voltage is gradually increased from 0 V, no current will flow until 0.3 V is exceeded. Once a current greater than or equal to 1 % of the range rating starts flowing, the current can flow at voltages less than equal to 0.3 V. *2 Condition in which the booster is connected to the master unit. *3 M range applies to the full scale of H range.