Changing the ENABLE signal activates the protection function. This action turns off the high voltage relay inside TOS5200 and stops the voltage output circuit. If current is flowing at this time, spike noise may occur, leading to voltage detection and causing the DANGER LED to light up briefly.
To avoid this, wait a few seconds after the test is completed before changing the ENABLE signal.