| Item | TOS9302 | TOS9303 | TOS9303LC |
|---|---|---|---|
| Current rise time settings range | 0.1 s to 200.0 s | ||
| Current fall time setting time *1 | 0.1 s to 200.0 s, OFF | ||
| Test time | 0.1 s to 1000.0 s, OFF | ||
| Accuracy | ±(100 ppm of setting + 20 ms) (excluding the fall time) | ||
*1 This setting is used only when a PASS judgment occurs. During a DC test, the voltage may not drop all the way within the set time because of the electrostatic capacity inside the product and the EUT.