PLZ164WL_Input /Output signal_J1 connector(26-pin MIL connector)
| PLZ164WL | PLZ334WL |
---|
Load on/off control input | Turn on the load with a high (or low) CMOS level signal |
---|
Load on status output | On when the load is on (open collector output from a photocoupler) |
---|
Range switch input | Switch ranges L, M, and H using a 2-bit signal |
---|
Range status output | Outputs range L, M, or H using a 2-bit signal (open collector output from a photocoupler) |
---|
Trigger input | Clear the sequence operation pause with a high CMOS level signal whose duration is 10 μs or longer |
---|
Alarm input | Activate the alarm with a low CMOS level signal |
---|
Alarm release input | Release the alarm with a low CMOS level signal |
---|
Alarm status output | On when OVP, OCP, OPP, OHP, UVP, or REV is activated or when an external alarm input is applied (open collector output from a photocoupler) |
---|
Short signal output | Relay contact output (30 Vdc/1 A) |
---|
External voltage control (CC, CR, and CP mode) | Voltages in the range of 0 V to 10 V correspond to 0 % to 100 % of the rated current (CC mode) or rated power (CP mode). Voltages in the range of 0 V to 10 V correspond to the range of resistance values from the maximum resistance value to the minimum resistance value (CR mode). |
---|
External voltage control (CV mode) | Voltages in the range of 0 V to 10 V correspond to the range of voltages from 0 % of the rated voltage to 100 % of the rated voltage. |
---|
External voltage control (superimposing in CC mode) | Superimpose the current on the CC mode panel/remote setting by applying an external voltage of -10 V to 10 V (CC mode). 0 V corresponds to 0 % of the current setting and 10 V corresponds to 100 % of the current setting. |
---|
Current monitor output | 10 V for f.s (H or L range), 1 V for f.s (M range) |
---|