PBZ_BP_Measurement function
| Item/Model | PBZ20-120 BP | PBZ20-140 BP | PBZ20-160 BP | PBZ20-180 BP | PBZ20-200 BP |
|---|
| Voltage measurement (DC) | Measurement range | 120 % of rating |
|---|
| Resolution | 0.001 V |
|---|
| Accuracy *1 | ± (0.05 % of rdng + 0.05 % of rtg) |
|---|
| Voltage measurement (AC and DC + AC) | Measurement range | AC | 120 % of rating/CF |
|---|
| DC+AC | 120 % of rating |
|---|
| Resolution | 0.001 V |
|---|
| Accuracy*1*2 | 5 Hz<f≦10 kHz | ±(0.5 % of reading + 0.1 % of rating) |
|---|
| 10 kHz<f≦50 kHz | ±(1 % of reading + 0.2 % of rating) |
|---|
| 50 kHz<f≦100 kHz | ±(2 % of reading + 0.2 % of rating) |
|---|
| Voltage measurement (PEAK) | Measurement range | 120 % of rating |
|---|
| Resolution | 0.01 V |
|---|
| Accuracy *1*3 | ±0.5 % of rating |
|---|
| Current measurement (DC) | Measurement range | 120 % of rating |
|---|
| Resolution | 0.006 A | 0.007 A | 0.008 A | 0.009 A | 0.010 A |
|---|
| Accuracy *1 | ±(0.3 % of reading+1.6 % of rating) | ±(0.3 % of reading+1.9 % of rating) | ±(0.3 % of reading+2.2 % of rating) | ±(0.3 % of reading+2.5 % of rating) | ±(0.3 % of reading+2.8 % of rating) |
|---|
| Temperature coefficient | ± (150 ppm / °C of rating) (TYP) |
|---|
| Current measurement (AC, DC + AC) | Measurement range | AC | 120 % of rating/CF |
|---|
| DC+AC | 120 % of rating |
|---|
| Resolution | 0.006 A | 0.007 A | 0.008 A | 0.009 A | 0.010 A |
|---|
| Accuracy*1*2 | 5 Hz<f≦10 kHz | ±(3 % of reading + 0.1 % of rating) |
|---|
| 10 kHz<f≦50 kHz | ±(10 % of reading + 1 % of rating) |
|---|
| Current measurement (PEAK) | Measurement range | 120 % of rating |
|---|
| Resolution | 0.06 A | 0.07 A | 0.08 A | 0.09 A | 0.10 A |
|---|
| Accuracy *1*3 | ±0.5 % of rating |
|---|
| Measurement time (Aperture) | 100 μs to 3600 s |
|---|
| Item/Model | PBZ40-60 BP | PBZ40-70 BP | PBZ40-80 BP | PBZ40-90 BP | PBZ40-100 BP |
|---|
| Voltage measurement (DC) | Measurement range | 120 % of rating |
|---|
| Resolution | 0.001 V |
|---|
| Accuracy *1 | ± (0.05 % of rdng + 0.05 % of rtg) |
|---|
| Voltage measurement (AC and DC + AC) | Measurement range | AC | 120 % of rating/CF |
|---|
| DC+AC | 120 % of rating |
|---|
| Resolution | 0.001 V |
|---|
| Accuracy*1*2 | 5 Hz<f≦10 kHz | ±(0.5 % of reading + 0.1 % of rating) |
|---|
| 10 kHz<f≦50 kHz | ±(1 % of reading + 0.2 % of rating) |
|---|
| 50 kHz<f≦100 kHz | ±(2 % of reading + 0.2 % of rating) |
|---|
| Voltage measurement (PEAK) | Measurement range | 120 % of rating |
|---|
| Resolution | 0.01 V |
|---|
| Accuracy *1*3 | ±0.5 % of rating |
|---|
| Current measurement (DC) | Measurement range | 120 % of rating |
|---|
| Resolution | 0.006 A | 0.007 A | 0.008 A | 0.009 A | 0.010 A |
|---|
| Accuracy *1 | ±(0.3 % of reading+1.6 % of rating) | ±(0.3 % of reading+1.9 % of rating) | ±(0.3 % of reading+2.2 % of rating) | ±(0.3 % of reading+2.5 % of rating) | ±(0.3 % of reading+2.8 % of rating) |
|---|
| Temperature coefficient | ± (150 ppm / °C of rating) (TYP) |
|---|
| Current measurement (AC, DC + AC) | Measurement range | AC | 120 % of rating/CF |
|---|
| DC+AC | 120 % of rating |
|---|
| Resolution | 0.006 A | 0.007 A | 0.008 A | 0.009 A | 0.010 A |
|---|
| Accuracy*1*2 | 5 Hz<f≦10 kHz | ±(3 % of reading + 0.1 % of rating) |
|---|
| 10 kHz<f≦50 kHz | ±(10 % of reading + 1 % of rating) |
|---|
| Current measurement (PEAK) | Measurement range | 120 % of rating |
|---|
| Resolution | 0.06 A | 0.07 A | 0.08 A | 0.09 A | 0.10 A |
|---|
| Accuracy *1*3 | ±0.5 % of rating |
|---|
| Measurement time (Aperture) | 100 μs to 3600 s |
|---|