OT03-PCR-WEA_General specifications

ItemSpecifications
Power supplyNominal input ratingSingle-phase 100 Vac to 240 Vac, 50 Hz to 60 Hz
Input voltage rangeSingle-phase 90 Vac to 250 Vac
Power consumption100 VA max
Environmental conditionsOperating environmentIndoor use, overvoltage category II
Operating temperature range0 °C to +50 °C (32 °F to 122 °F)
Operating humidity range20 %rh to 80 %rh(no condensation)
Storage temperature range-10 °C to +60 °C (14 °F to 140 °F)
Storage humidity range90 %rh or less (no condensation)
AltitudeUp to 2000 m
Cooling methodForced air cooling using fan
Insulation resistanceBetween inlet input and chassis
Between inlet input and input/output, Between input/output and chassis
500 Vdc, 30 MΩ or more
Withstanding voltageBetween inlet input and chassis
Between inlet input and chassis, Between inlet input and input/output
No abnormalities at 1500 Vac for 1 minute.
Between input/output and chassisNo abnormalities at 1600 Vac for 1 minute.
Earth continuityAC 25 A, 0.1 Ω or less
Input terminal blockM5 terminal block x 7P
Output terminal blockM5 terminal block x 7P
Electromagnetic compatibility *1*2Complies with the requirements of the following directive and standard.
  • EMC Directive 2014/30/EU
  • EN 61326-1(Class A) *3 )
  • EN 55011(Class A) *3, Group 1 *4 )
  • EN 61000-3-2
  • EN 61000-3-3
Applicable under the following conditions:
The maximum length of all cabling and wiring connected to the product must be less than 3 m.
SafetyComplies with the requirements of the following directive and standard.
Low Voltage Directive 2014/35/EU *2 *2
EN 61010-1(Class I) *5, Pollution degree 2 *6 )

*1 Not applicable to custom-ordered or modified products. *2 Only for models with CE marking/UKCA marking displayed on the main body. *3 This product is Class A equipment. It is intended for use in an industrial environment. Use of this product in a residential area may cause interference. In such cases, special measures to reduce electromagnetic emissions by the user may be required to prevent interference with radio and television broadcast reception. *4 This product is Group 1 equipment. This product does not intentionally generate/use radio frequency energy in the form of electromagnetic emissions, induction and/or electrostatic coupling for material processing or inspection/analysis. *5 This product is Class I equipment. The protective conductor terminals of this product must be grounded. Safety is not guaranteed if not properly grounded. *6 Contamination is a condition in which foreign matter (solid, liquid, or gas) adheres to the product which may cause a decrease in Insulation resistance or surface resistivity. Contamination Dye degree 2 assumes that only non-conductive contamination is present, which can sometimes become temporarily conductive due to condensation.