Item | Specifications | |
---|---|---|
Power supply | Nominal input rating | Single-phase 100 Vac to 240 Vac, 50 Hz to 60 Hz |
Input voltage range | Single-phase 90 Vac to 250 Vac | |
Power consumption | 100 VA max | |
Environmental conditions | Operating environment | Indoor use, overvoltage category II |
Operating temperature range | 0 °C to +50 °C (32 °F to 122 °F) | |
Operating humidity range | 20 %rh to 80 %rh(no condensation) | |
Storage temperature range | -10 °C to +60 °C (14 °F to 140 °F) | |
Storage humidity range | 90 %rh or less (no condensation) | |
Altitude | Up to 2000 m | |
Cooling method | Forced air cooling using fan | |
Insulation resistance | Between inlet input and chassis
Between inlet input and input/output, Between input/output and chassis | 500 Vdc, 30 MΩ or more |
Withstanding voltage | Between inlet input and chassis
Between inlet input and chassis, Between inlet input and input/output | No abnormalities at 1500 Vac for 1 minute. |
Between input/output and chassis | No abnormalities at 1600 Vac for 1 minute. | |
Earth continuity | AC 25 A, 0.1 Ω or less | |
Input terminal block | M5 terminal block x 7P | |
Output terminal block | M5 terminal block x 7P | |
Electromagnetic compatibility *1*2 | Complies with the requirements of the following directive and standard.
The maximum length of all cabling and wiring connected to the product must be less than 3 m. | |
Safety | Complies with the requirements of the following directive and standard.
Low Voltage Directive 2014/35/EU *2 *2 EN 61010-1(Class I) *5, Pollution degree 2 *6 ) |
*1 Not applicable to custom-ordered or modified products. *2 Only for models with CE marking/UKCA marking displayed on the main body. *3 This product is Class A equipment. It is intended for use in an industrial environment. Use of this product in a residential area may cause interference. In such cases, special measures to reduce electromagnetic emissions by the user may be required to prevent interference with radio and television broadcast reception. *4 This product is Group 1 equipment. This product does not intentionally generate/use radio frequency energy in the form of electromagnetic emissions, induction and/or electrostatic coupling for material processing or inspection/analysis. *5 This product is Class I equipment. The protective conductor terminals of this product must be grounded. Safety is not guaranteed if not properly grounded. *6 Contamination is a condition in which foreign matter (solid, liquid, or gas) adheres to the product which may cause a decrease in Insulation resistance or surface resistivity. Contamination Dye degree 2 assumes that only non-conductive contamination is present, which can sometimes become temporarily conductive due to condensation.