Dimensions / Weight
W | H | D | |
---|---|---|---|
Dimensions | 214 mm (8.43 inch) |
80 mm (3.15 inch) |
300 mm (11.81 inch) |
Battery Impedance Meter
BIM1000 Series
BIM1000 series
260,000円 – 420,000円
- Maximum voltage measurement: 1000V max. (BIM1100), 300V max. (BIM1030)
- Voltage measurement accuracy: ±(0.01% of reading +3 digit)
- Resistance measurement accuracy: ±(0.5% of reading +5 digit)
- Resistance measurement ranges: 3mΩ / 30mΩ / 300mΩ / 3Ω
- High resolution: Voltage 10μV (6V range), Resistance 0.1μΩ (3mΩ range)
- Measurement frequency: 1kHz ±0.2Hz
- Sampling speed (Resistance & voltage simultaneous measure-ments): 20ms (FAST)
- Zero Adjustment Function: Effective for decreasing measure-ment error.
- Measurement logging (500 pairs) and collective transfer function
- SIGNAL I/O, RS232C and USB as standard interface (BIM1030 WITH LAN and BIM1100 WITH LAN are standard equipped also with a LAN port)
- New high visibility color display.
Line-up
Image | SKU | 標準価格 | WEBかんたんお見積り | ||
---|---|---|---|---|---|
BIM1030 | ±300 V | with LAN | 260,000円 (税込:286,000円) | ||
BIM1030 with LAN | ±300 V | with LAN | 280,000円 (税込:308,000円) | ||
BIM1100 | ±1000 V | with LAN | 400,000円 (税込:440,000円) | ||
BIM1000 series | ±1000 V | with LAN | 420,000円 (税込:462,000円) |
Overview
An ideal unit for manufacturing and inspection applications of rechargeable batteries.
Ever-changing battery technology requires batteries powering electric vehicles to have high voltage, high power and low impedance. The Battery Impedance Meter, or BIM1000 Series, is capable of measuring up to 1000 V of test voltage for simultaneous measurements of both battery voltage and resistance at high speeds. The BIM is the ideal equipment for power battery development research and production tests.
Color liquid crystal display (LCD) / Comparator Functions
High visibility color monitor. The resistance, voltage, upper and lower limits values are displayed at a glance.
The comparator functions allows setting HIGH/LOW, and resistance and voltage can be simultaneously judged by independent comparators. Judgment results are provided on the display. External I/O is available for signal output. Applications
Specifications
Voltmeter (The range can be AUTO setting available)
Item | BIM1030 / BIM1100 | ||||
---|---|---|---|---|---|
Rated input | BIM1030 : ± 300V / BIM1100 : ± 1000V | ||||
Range | 6V | 60V | 300V(BIM1030) | 600V(BIM1100) | 1000V(BIM1100) |
Maximum display value *1 | ±6.30000V | ±63.0000V | ±315.000V | ± 630.000V | ±1050.00V |
Resolution | 10μV | 100μV | 1mV | 10mV | |
Accuracy *2 | ± ( 0.01% of reading + 3digit ) | ||||
Temperature coefficient | ± ( 0.001% of reading + 0.3digit ) / ℃ | ||||
Response time *3 | Approx. 1 ms |
*1. Displays OVER when the measurement range is exceeded. *2. Add ±2 digit when the sampling speed is set to FAST or MID. *3. The time for the product’s measurement circuit to stabilize when a probe in an open state is connected to the DUT.
Resistance meter (The range can be AUTO setting available)
Item | BIM1030 / BIM1100 | |||
---|---|---|---|---|
Measurement method | Four-terminal measurement method | |||
Range | 3mΩ | 30mΩ | 300mΩ | 3Ω |
Maximum display value *1 | 3.1000mΩ | 31.000mΩ | 310.00mΩ | 3.1000Ω |
Resolution | 0.1μΩ | 1μΩ | 10μΩ | 100μΩ |
Measured current *2 | 100mA | 10mA | 1mA | |
Measurement frequency | 1kHz ± 0.2Hz | |||
Accuracy *3 | ±( 0.5% of reading + 5digit) | |||
Temperature coefficient | ± ( 0.05% of reading + 1digit ) / ℃ | ± ( 0.05% of reading + 0.5digit) / ℃ | ||
Response time *4 | Approx. 2 ms |
*1. Displays OVER when the measurement range is exceeded. *2. Within error ± 10 %. *3. Add ±3 digit when the sampling speed is set to FAST and ±2 digit when the sampling speed is set to MID. *4. The time for the product’s measurement circuit to stabilize when a probe in an open state is Environ- ment Power supply Spec guaranteed range Operating range Storage range Temperature Humidity Temperature Humidity Temperature Humidity connected to the DUT.
Sampling time
Item | BIM1030 / BIM1100 | ||||
---|---|---|---|---|---|
Sampling speed | FAST | MEDIUM | SLOW | ||
Power supply frequency | 50Hz | 20ms | 50ms | 160ms | |
±60Hz | 42ms | 150ms |
Judgment feature
Item | BIM1030 | BIM1100 | |
---|---|---|---|
Judgment method | Window comparator method. Judgment made with software. | ||
Resistance | Setting range | 0.0001Ω 〜 3.1000Ω | |
Resolution | 100μΩ | ||
Voltage | Setting range | 0.000V 〜 315.000V | 0.00V 〜 1050.00V |
Resolution | 1mV | 10mV |
Interface
Item | BIM1030 / BIM1100 |
---|---|
RS232C | D-SUB 9-pin connector, EIA-232-D compliant |
USB | Complies with USB Specification 2.0. Data rate: 12 Mbps max. (Full Speed) |
Complies with USBTMC Specification 1.0 and USBTMC-USB488 Specification 1.0 | |
LAN *1 | IEEE 802.3, 100Base-Tx/10Base-T Ethernet, IPv4, RJ45, Auto-negotiation supported but Auto MDI/MDI-X not supported. |
SIGNAL I/O | D-SUB 25-pin connector. |
*1. BIM1030 WITH LAN or BIM1100 WITH LAN only
General specifications
Item | BIM1030 | BIM1100 | ||
---|---|---|---|---|
Environment | Installation location | Indoors, 2000 m or less | ||
Spec guaranteed range | Temperature / Humidity | 18 °C to 28 °C (-4 °F to 158 °F) / 20 %rh to 80 %rh (no condensation) | ||
Operating range | Temperature / Humidity | 0 °C to 40 °C (32 °F to +122 °F) / 20 %rh to 80 %rh (no condensation) | ||
Storage range | Temperature / Humidity | -10 °C to 60 °C (-4 °F to 158 °F) / 90 %rh or less (no condensation) | ||
Power supply | Input voltage range / Input frequency range | 85 Vac to 264 Vac (100 Vac to 240 Vac)/47 Hz to 63 Hz | ||
Rated power | 30VA | |||
Isolation voltage | ± 300 V max | ± 1000 V max | ||
Insulation resistance | 30 MΩ or more (500 Vdc)(between AC LINE and chassis) | |||
Withstanding voltage | Between the AC LINE and the chassis | 1500 Vac for 1 minute, 10 mA or less | ||
Between all the mea- surement terminals and the chassis | 2000 Vdc for 1 minute, 1 mA or less | |||
Between all the mea- surement terminals and SIGNAL I/O | 2000 Vdc for 1 minute, 1 mA or less | |||
External dimensions / Weight | 214(8.43)W×80(3.15)H×300(11.81)D mm(inches) (Does not include protrusions)/ Approx. 3 kg (6.6 lbs) | |||
Electromagnetic compatibility (EMC) *1 *2 | Complies with the requirements of the following directive and standards. EMC Directive 2014/30/EU EN 61326-1 (Class A *3), EN 55011 (Class A *3, Group 1 *4), EN 61000-3-2, EN 61000-3-3 | |||
Safety *1 | Complies with the requirements of the following directive and standards. Low Voltage Directive 2014/35/EU *2 EN 61010-1 (Class I *5, Pollution Degree 2 *6), EN 61010-2-030, EN 61010-031 |
*1. Does not apply to specially ordered or modified products. *2. Only on models that have CE/UKCA marking on the panel. *3. This is a Class A instrument. This product is intended for use in an industrial environment. This prod- uct may cause interference if used in residential areas. Such use must be avoided unless the user takes special measures to reduce electromagnetic emissions to prevent interference to the reception of radio and television broadcasts. *4. This is a Group 1 instrument. This product does not generate and/or use intentionally radio-frequen- cy energy, in the form of electromagnetic radiation, inductive and/or capacitive coupling, for the treatment of material or inspection/analysis purpose. *5. This is a Class I instrument. Be sure to ground this product’s protective conductor terminal.The safety of this product is guaranteed only when the product is properly grounded. *6. Pollution is addition of foreign matter (solid, liquid or gaseous) that may produce a reduction of dielectric strength or surface resistivity. Pollution Degree 2 assumes that only non-conductive pollu- tion will occur except for an occasional temporary conductivity caused by condensation.