High-performance multifunctional AC Power Supplies

PCR-LE2 Series

PCR-LE2-SERIES

3,700,00014,000,000

  • High-quality/high-stability output with a high-speed linear amp
  • Capable of various power line abnormality simulations and sequence operation
  • Single-phase 500 VA to 9 kVA, supporting the system for the single-phase, and expandable with optional drivers for the single-phase three-wire, and three-phase operation
  • Expandable capacity up to 27 kVA (single-phase), 54 kVA (single-phase three-wires), and 81 kVA (three-phase)
  • Equipped with various measuring functions
  • Features a full range of measuring functions and supports AC, DC, and AC + DC Outputs
  • Detachable front panel
  • Eco-friendly function equipped

For PCR12000LE2, PCR18000LE2, and PCR27000LE2, installation costs are required separately.


エミッション試験 IEC 61000-3-11(要 フリッカ測定器)
アビオニクス
Login to myKIKUSUI / What is myKIKUSUI?
ImageSKU入力:皮相電力(kVA以下)Output rating AC mode: Power capacity標準価格WEBかんたんお見積り
PCR-LE2シリーズPCR6000LE210.66kVA3,700,000
(税込:4,070,000
PCR-LE2シリーズPCR9000LE215.79kVA5,800,000
(税込:6,380,000
PCR-LE2シリーズPCR12000LE22312kVA6,980,000
(税込:7,678,000
PCR-LE2シリーズPCR18000LE23318kVA10,000,000
(税込:11,000,000
PCR-LE2シリーズPCR27000LE24827kVA14,000,000
(税込:15,400,000
SKU: PCR-LE2-SERIES

Line-up

ZOOM
ModelOutput capacityMaximum output currentAC mode (L/H range)DC mode (L/H range)Price
 Single-phase, Three phase 4-wireSingle phase 3-wireSingle phaseSingle phase 3-wire Single phaseThree phase 4-wire Single phaseSingle phase 3-wire 
PCR6000LE26kVA4kVA60A/30A20A/10A1 V to 150 V / 2 V to 300 V60A/30A20A/10A±1.4 V to ±212 V / ±2.8 V to ±424 V42A/21A14A/7A3,700,000 (税込:4,070,000
PCR9000LE29kVA6kVA90A/45A30A/15A90A/45A30A/15A63A/31.5A21A/10.5A5,800,000 (税込:6,380,000
PCR12000LE212kVA8kVA120A/60A40A/20A120A/60A40A/20A84A/42A28A/14A6,980,000 (税込:7,678,000
PCR18000LE218kVA12kVA180A/90A60A/30A180A/90A60A/30A126A/63A42A/21A10,000,000 (税込:11,000,000
PCR27000LE227kVA18kVA270A/135A90A/45A270A/135A90A/45A189A/94.5A63A/31.5A14,000,000 (税込:15,400,000

Specifications

Input ratings (AC rms)

ZOOM
Item/ModelPCR6000LE2PCR9000LE2PCR12000LE2PCR18000LE2PCR27000LE2
3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V
VoltageLine voltage 170 V to 250 VLine voltage 324 V to 440 V (Phase voltage 187 V to 254 V)Line voltage 170 V to 250 VLine voltage 324 V to 440 V (Phase voltage 187 V to 254 V)Line voltage 170 V to 250 VLine voltage 324 V to 440 V (Phase voltage 187 V to 254 V)Line voltage 170 V to 250 VLine voltage 324 V to 440 V (Phase voltage 187 V to 254 V)Line voltage 170 V to 250 VLine voltage 324 V to 440 V (Phase voltage 187 V to 254 V)
PhaseSingle phaseThree phase 3-wireThree phase 4-wireThree phase 3-wireThree phase 4-wireThree phase 3-wireThree phase 4-wireThree phase 3-wireThree phase 4-wireThree phase 3-wireThree phase 4-wire
Frequency47 Hz to 63 Hz
Apparent powerApprox. 10.6 kVAApprox. 15.7 kVAApprox. 23 kVAApprox. 33 kVAApprox. 48 kVA
Power factor *10.97 (TYP)
Max. current64 A or less38A or less21A or less55A or less30A or less75A or less39A or less111A or less59A or less165A or less91A or less

*1 When the output phase voltage is 100 V or 200 V, the output current is the rated value, the load power factor is 1, and the output frequency is between 40 Hz and 999.9 Hz.


AC mode output ratings (AC rms)

ZOOM
Item/ModelPCR6000LE2PCR9000LE2PCR12000LE2PCR18000LE2PCR27000LE2
3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V
Phase voltage (output L range, output H range)*21 V to 150 V / 2 V to 300 V
Voltage setting range0 V to 152.5 V / 0 V to 305.0 V
Voltage setting accuracy (output L range, output H range)*3±(0.3 % of set + 0.6 V)
Max. current*4Single phase, poly phase, L range, H range60 A, 30 A · 20 A, 10 A90 A, 45 A · 30 A, 15 A120 A, 60 A · 40 A, 20 A180 A, 90 A · 60 A, 30 A270 A, 135 A · 90 A, 45 A
Phase*5Single phase · Single phase3-wire · Three phase 4-wire
Power capacitySingle phase, Three-phase 4-wire, Single phase 3-wire6 kVA · 4 kVA9 kVA · 6 kVA12 kVA · 8 kVA18 kVA · 12 kVA27 kVA · 18 kVA
Maximum peak current*6Max. current (rms) × 4 (TYP)
Max. reverse current*730 % of the max. current (rms)
Load power factor*40 to 1 (leading or lagging)
Frequency*4 *8 *91 Hz to 999.9 Hz

*2 L/H range can be changed by means of a switch on the front panel. Resolution: 0.1V
*3 出When the output frequency is between 45 Hz and 65 Hz, with no load, and at room temperature. *4 When the maximum voltage is between 1 V and 100 V (L range) or 2 V and 200 V (H range) and the load power factor is between 0.8 and 1.When the output phase voltage is between 100 V and 150 V or 200 V and 300 V (AC mode) or ±100 V and ±212 V or ±200 V and ±424 V (DC mode), the output current is reduced by the output phase voltage.
When the load power factor is between 0 and 0.8, the output current is reduced by the load power factor. (AC mode) When the output frequency is between 1 Hz and 40 Hz, the output current is reduced by the output frequency. (AC mode) *5 The output phase mode can be changed by means of a key on the operation panel. “Poly” in the table indicates single- phase three-wire mode and three-phase four-wire mode. *6 When the output phase voltage is in the vicinity of the peak (±15 deg) (However, this is limited by the rated output current’s rms value). *7 When the output phase voltage is 100 V or 200 V and the output frequency is between 40 Hz and 999.9 Hz (reverse current is −90 deg to −180 deg / 90 deg to 180 deg out of phase with the output voltage).
*8 Resolution : 0.01Hz (1.00 Hz to 100.0 Hz), 0.1Hz(100.0 Hz to 999.9 Hz) *9 The “500Hz Limit Model” limits the maximum frequency up to 500Hz under the “Three-phase output”.
PCR-LE2 Series 500Hz Limit Model
The PCR-LE Series offers the type on each model that limits the maximum output frequency up to 500 Hz.


DC mode output ratings, AC+DC mode (for Single-phase and Single-phase Three-wire output only)

ZOOM
Item/ModelPCR6000LE2PCR9000LE2PCR12000LE2PCR18000LE2PCR27000LE2
3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V
Phase voltage (output L range, output H range)*2±1.4 V to ±212 V / ±2.8 V to ±424 V
Voltage setting range-215.5 V to +215.5 V / -431.0 V to +431.0 V
Voltage setting accuracy (output L range, output H range) *10± (0.05 % of set + 0.05 V / 0.1 V)
Max. current*4Single phase, Single phase 3-wire andThree-phase, L range, H range42 A, 21 A · 14 A, 7 A63 A, 31.5 A · 21 A, 10.5 A84 A, 42 A · 28 A, 14 A126 A, 63 A · 42 A, 21 A189 A, 94.5 A · 63 A, 31.5 A
Max. instantaneous current*11Max. current (rms) × 3.6
Power capacitySingle phase, Single phase 3-wire, Three-phase4.2 kW · 2.8 kW6.3 kW · 4.2 kW8.4 kW · 5.6 kW12.6 kW · 8.4 kW18.9 kW · 12.6 kW

*2 L/H range can be changed by means of a switch on the front panel. Resolution: 0.1V
*4 When the maximum voltage is between 1 V and 100 V (L range) or 2 V and 200 V (H range) and the load power factor is between 0.8 and 1.When the output phase voltage is between 100 V and 150 V or 200 V and 300 V (AC mode) or ±100 V and ±212 V or ±200 V and ±424 V (DC mode), the output current is reduced by the output phase voltage.
When the load power factor is between 0 and 0.8, the output current is reduced by the load power factor. (AC mode) When the output frequency is between 1 Hz and 40 Hz, the output current is reduced by the output frequency. (AC mode) *10 With no load at room temperature
*11 Limited by the rated output current’s rms value


Output voltage stability

ZOOM
Item/ModelPCR6000LE2PCR9000LE2PCR12000LE2PCR18000LE2PCR27000LE2
3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V
Line regulation(With respect to changes in the rated range)Within ±0.1 %
Line regulation(With respect to 0 % to 100 % changes in the rating)*12±0.3 V±0.5 V
Output frequency variation in AC mode(Between 40 Hz and 999.9 Hz)*13Within ±0.5 %Within ±1 %
Ripple noise in DC mode(5 Hz to 1 MHz components)0.25 Vrms or less0.5 Vrms or less
Ambient temperature variation(With respect to changes in the rated range)*14100 ppm/ °C (TYP)

*12 When the output phase voltage is between 80 V and 150 V (L range) or 160 V and 300 V (H range) and the load power factor is 1. At the output terminal block. When the response mode is set to MEDIUM.(There is no F mode)
*13 When the output phase voltage is between 80 V and 150 V (L range) or 160 V and 300 V (H range) and the load power factor is 1. This is the output line regulation with 200 Hz as the reference. When the response mode is set to MEDIUM.(There is no F mode)
*14 When the output phase voltage is 100 V or 200 V and the output current is 0 A.


Output frequency stability, output voltage waveform distortion ratio, output voltage response speed, efficiency

ZOOM
Item/ModelPCR6000LE2PCR9000LE2PCR12000LE2PCR18000LE2PCR27000LE2
3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V
Output frequency stability(With respect to changes in all rated ranges)Within ±5×10– 5, Setting accuracy : Within ±1×10– 4
Output voltage waveform distortion ratio*150.3 % or less0.5% or less
Output voltage response speed*1630 μs (TYP)50μs (TYP)
Efficiency*158 % or more
Phase difference of the output phase voltage *17Resolution1deg
AccuracyWithin ± (0.4° + f0×1.8×10– 3) deg f0 is the output frequency *18

*1 When the output phase voltage is 100 V or 200 V, the output current is the rated value, the load power factor is 1, and the output frequency is between 40 Hz and 999.9 Hz. *15 When the output phase voltage is between 80 V and 150 V (L range) or 160 V and 300 V (H range) and the load power factor is 1. When the response mode is set to MEDIUM.(There is no F mode) *16 When the output phase voltage is 100 V or 200 V, the load power factor is 1, and the output current changes from 0 A to the rated value and from the rated value to 0 A. *17 Phase difference between output voltages (phase voltages) when each phase is considered along with the neutral point.


Meters (fluorescent display)

ZOOM
Item/ModelPCR6000LE2PCR9000LE2PCR12000LE2PCR18000LE2PCR27000LE2
3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V
Voltmeter *19 *20ResolutionRMS,AVE Display mode0.1 V
AccuracyRMS,AVE Display modeWithin ± (1 % of rdng + 2 digits) (10 V to 848 V and at room temperature)
Ammeter *19 *20ResolutionRMS,AVE Display mode Single phase · Poly phase0.1A · 0.01 A0.1 A0.1 A / 1 A · 0.1 A
AccuracyRMS Display modeWithin ± (1% of reading + 2digits) (5 % of the max. rated current to max. rated current and at room temperature)
Wattmeter*20ResolutionSingle phase · Poly phase1 W · 0.1 W / 1 W1 W1 W / 10 W
Accuracy Within ± (1 % of reading + 3digits) (10 % of the rated power capacity to the rated power capacity, when the load power factor is 1, and at room temperature.)
Frequency meter*21Resolution0.01 Hz / 0.1 Hz

*19 With the true rms display, a waveform with a crest factor of 3 or less.
*20 When the output frequency is between 45 Hz and 65 Hz.
*21 Displays the output frequency setting (frequency of the internal reference voltage)


General

ZOOM
Item/ModelPCR6000LE2PCR9000LE2PCR12000LE2PCR18000LE2PCR27000LE2
 3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V3P3W200V3P4W400V
Insulation resistanceBetween input and chassis, output and chassis, input and output500 Vdc, 10 MΩ or more
Withstanding voltageBetween input and chassis, output and chassis, input and output1.5 kVac for 1 minute
Circuit methodLinear amplifier system
Environmental conditionsOperating temperature range / Storage temperature range0 °C to +50 °C / -10 °C to +60 °C
Operating humidity range / Storage humidity range20 % rh to 80 % rh (no condensation) / 90 % rh or less (no condensation)
WeightApprox.140 kg (308.6 Ibs)Approx.190kg (418.8 Ibs)Approx.350 kg (771.6 Ibs)Approx.480 kg (1058.2 Ibs)Approx.630 kg (1388.9 Ibs)
Input terminalInput terminal board [3φ]M8M5M5M8M8M8
Output terminalOutput terminal board Single phase · Single phase 3-wire,Three-phase 4-wireM8 · M5M8 · M8
Input power cord [Sold separately option]Typesingle-core cableRequired for the installation work, contact local distributor.
The number3 pc4 pc5 pc4 pc5 pc
Conductor cross section/Length14mm2/3m8mm2/3m5.5mm2/3m14mm2/3m5.5mm2/3m
AccessoriesSetup guide1 copy
Quick Reference1 each for English and Japanese
Safety information1 copy
CD-ROM(User’s manual)1 disc
Heavy Duty Warning Seals1 disc
OtherElectromagnetic compatibility (EMC)EMC Directive 2014/30/EU. EN61326-1. EN 55011 The maximum length of all cables and wires connected to the PCR-LE Series must be less than 3 m.
SafetyLow Voltage Directive 2014/35/EU, EN61010-1 (Class I, Pollution Degree2)IEC 61010-1 (Class I, Pollution Degree2)

Limit Values and Protection Functions

ZOOM
Item/ModelPCR6000LE2PCR9000LE2PCR12000LE2PCR18000LE2PCR27000LE2
VoltageAC voltage upper limit AC voltage lower limit0.0 V to 305.0 V
DC voltage upper limit DC voltage lower limit-431.0 V to +431.0 V
Output overvoltage protection AC/AC+DC mode0.0 V to 474.1 V
Output overvoltage protection DC mode-474.1 V to +474.1 V
Output undervoltage protection AC/AC+DC mode0.0 V to 474.1 V
Output undervoltage protection DC mode-474.1 V to +474.1 V
Resolution0.1 V
FrequencyUpper limit Lowerlimit1 Hz to 999.9 Hz, 500 Hz LMT model: 1 Hz to 500 Hz (Three-phase output)
Resolution0.01 Hz (1.00 Hz to 100.0 Hz), 0.1 Hz (100.0 Hz to 999.9 Hz)
CurrentCurrent limit *1 AC modeSingle-phase output6.00A to 66.00A9.00A to 99.00A12.00A to 132.0A18.00A to 198.0A27.00A to 297.0A
Single-phase three-wire output Three-phase output2.00A to 22.00A3.00A to 33.00A4.00A to 44.00A6.00A to 66.00A9.00A to 99.00A
Current limit *1 DC/AC+DC modeSingle-phase output4.20A to 46.20A6.30A to 69.30A8.40A to 92.40A12.60A to 138.6A18.90A to 207.9A
Single-phase three-wire output Three-phase output1.40A to 15.40A2.10A to 23.10A2.80A to 30.80A4.20A to 46.20A6.30A to 69.30A
Positive peak current limit *2Single-phase output6.00A to 264.0A9.00A to 396.0A12.00A to 528.0A18.00A to 792.0A27.00A to 1188A
Single-phase three-wire output Three-phase output2.00A to 88.00A3.00A to 132.0A4.00A to 176.0A6.00A to 264.0A9.00A to 396.0A
Negative peak current limit *2Single-phase output-6.00A to -264.0A-9.00 A to -396.0 A-12.00 A to -528.0 A-18.00 A to -792.0 A-27.00 A to -1188 A
Single-phase three-wire output Three-phase output-2.00A to -88.00A-3.00 A to -132.0 A-4.00 A to -176.0 A-6.00A to -264.0A-9.00 A to -396.0 A
Resolution *30.01 A (0.35 A to 100.0 A) , 0.1A (100.0 A to 1000 A) , 1 A (1000 A to 1188 A)

*1 The current that can actually be supplied is 1.1 times the rated current or the current limit, whichever is less. *2 The current that can actually be supplied is the maximum peak current or the current limit, whichever is less. *3 You can set the current in 0.01 A/ 0.1 A/ 1 A steps, but it may not change at this resolution depending on the relationship with the internal D/A resolution.


Options

Input power cable

For PCR6000LE2

For PCR6000LE2(3P3W input)/PCR9000LE2(3P3W input)

For PCR6000LE2(3P4W input)/PCR9000LE2(3P4W input)


Interface

For GPIB

for USB

For LAN/LXI

Analog signal interface boards※3 ※4 ※5 ※6

IEC Dip·Simulator

Line Impedance Network

Extension cable for control panel

Base Hold Angles

1 PCR500LE and PCR1000LE cannot be used. 2 Up to 5 units can be used in parallel. (Power within 27kVA) *3 Input waveforms are amplified and output as they are. 4 Multi-output models (6000LE2, 9000LE2) cannot amplify input waveforms in multi-phase output mode. 5 The output AC voltage value can be varied according to the input DC signal. 6 The multi-output models (6000LE2, 9000LE2) cannot control the amplitude in the multi-phase output mode.

2D CAD

PCR6000LE2

Outline | PDF 105.86 KB
Icon
Outline | DXF 251.83 KB
Icon

    PCR9000LE2

    Outline | PDF 96.04 KB
    Icon
    Outline | DXF 284.93 KB
    Icon

      PCR12000LE2

      Outline | PDF 383.00 KB
      Icon
      Outline | DXF 543.09 KB
      Icon

        PCR18000LE2

        Outline | PDF 426.49 KB
        Icon
        Outline | DXF 659.61 KB
        Icon

          PCR27000LE2

          Outline | PDF 314.28 KB
          Icon
          Outline | DXF 672.55 KB
          Icon

            Operation Manuals

            PCR_LE2
            Icon

              Please feel free to contact us with any questions or concerns.

              Download a Product Catalog View List
              Request a Quotation Contact Us
              close

              Search