Fixed the problem that a different network is set when the power is restarted after performing a specific operation in the meter mode (Meter) of the leakage current test (LC).
Firmware Version 1.18
2022-9-06
IFC1.08.0138/IOC1.08.0112/FPGA1.04.0007
Improvement
Improved the stability at long-term operation.
Resolved issues
Fixed the character display.
Firmware Version 1.17
2020-8-19
FPGA1.04.0007/IOC1.07.0105/IFC1.07.0133
Improvement
Improved the UI for Memory.
Improvement
Improved the peak accuracy in LC.
Improvement
Improved the over range processing of the Meter mode.
Improvement
Improved the behavior to reduce the operation load of FPGA.
Firmware Version 1.15
2020-5-19
FPGA1.03.0006/IOC1.06.0100/IFC1.06.0131
Resolved issues
Fixed the problem that the Interlock and the Calibration were not working at the same time when it was set to ‘Protect’ at start-up.
Resolved issues
Fixed the character display.
Resolved issues
Fixed the problem that the operation is unstable when starting the ‘AUTO’ test by entering the command from the Memory screen at Local mode.
Updated
Changed not to allow the Test Level changing during the Faill Time.
Improvement
Improved the behavior at the Contack Check of EC.
Improvement
Improved the stability at long-term operation.
Firmware Version 1.14
2020-1-17
FPGA1.02.0005/IOC1.05.0093/IFC1.05.0130
Resolved issues
Fixed the problem that the response of auto test step number for 'RES:REM?' query was always '1'.
Resolved issues
Fixed the problem that C-FAIL was saved as L-FAIL in Memory.
Resolved issues
Fixed the problem that Calibration Protection was not working at power-on.
Resolved issues
Fixed the problem that the previous test condition was recalled at one time when the auto test memory was recalled through the signal I/O interface.
Improvement
Improved the behavior when 'IP Address Method' is set to 'Static'.
Improvement
Improved the firmware updater to display its version on the panel.
Improvement
Increased the memory items saved in Memory.
Improvement
Displays 'CHECK' on the panel during checking the scanner contact (Contact Check).
Updated
Changed the panel display from 'error' to 'over' when the measurement result exceeds the measurement range.
Updated
Changed the maximum period of date setting to 2037.
Updated
Changed not to allow the Test Level key pressing during the rise time.
Resolved issues
Fixed the problem that Calibration Protection was released by pressing STOP key even it was set to 'Enable'.
Firmware Version 1.12
2019-7-19
IFC1.03.0123/IOC1.04.0077/FPGA1.02.0005
Resolved issues
Fixed the problem that was preventing tests from starting in rare cases when using the HP01A/02A-TOS.
Resolved issues
Fixed the problem that was preventing the next step from being performed in rare cases in an AUTO test even when FAIL Judgment was set to continue.
Resolved issues
Fixed the problem that was causing the measurement screen to display an error when an AUTO test was performed after a single test.
Resolved issues
Fixed the problem that was preventing an ACW test from being performed in rare cases even when a test was started by specifying an IR or ACW test.
Improvement
Improved the flickering measurement data display.
Resolved issues
Fixed the problem that was causing the instrument to return to READY immediately after a U-Fail occurrence in an AUTO test.
Resolved issues
Fixed the problem that was preventing the protection function from being activated even in an event of an overload during rising.
Resolved issues
Fixed the problem that was causing Over Heat protection and Power Supply protection to be misjudged in rare cases when long-term operation was performed.
Resolved issues
Fixed the problem that was causing the STOP bit to be set in the OPER:TEST register even when STOP was not pressed.
Resolved issues
Fixed the problem that was preventing the voltage rise rate judgment function from operating normally in rare cases.
Resolved issues
Fixed the problem that was causing L-FAIL to occur in rare cases when an IR test was started with the output open.
Resolved issues
Fixed the problem that was causing the instrument to operate at 1/10 of the specified upper and lower judgment values in an IR test during an auto test.
Resolved issues
Fixed the problem that was preventing tests from being performed according to the steps when a test was re-executed after releasing a protection function that was activated during an auto test.
Improvement
Improved the internal communication reliability.
Firmware Version 1.10
2019-5-08
IFC1.07.0107
Improvement
Changed the digits and units displayed during measurement * Only the display was changed (the accuracy was not changed).
Resolved issues
Changed the digits and units displayed during measurement * Only the display was changed (the accuracy was not changed).
Resolved issues
The test time was set to 1 second. Fixed the problem that was causing the test time to be set to 0.2 seconds when the power was restarted.
Resolved issues
Fixed the problem that was preventing auto tests from completing when pass judgments were made (the problem kept repeating the test).
Resolved issues
Fixed the problem that was causing unused bits to be returned in response to the STAT:OPER:TEST:COND? command.
Improvement
Fixed the problem that was causing errors with the FUNC PROG command.
Firmware Version 1.01
2019-1-29
IFC1.00.0092
Resolved issues
Fixed the problem that was preventing the instrument from starting during discharge.
Please feel free to contact us with any questions or concerns.
Cookies will be used to track which website is being visited. On our website, we certify the information that can be accessed by visitors using cookies. We also track viewed information by users to understand your interest and provide you with a better content. You will be able to disable cookie itself, you may choose the way to use personal identification whenever you use service that requires authentication.
Strictly Necessary Cookies
Strictly Necessary Cookie should be enabled at all times so that we can save your preferences for cookie settings.
If you disable this cookie, we will not be able to save your preferences. This means that every time you visit this website you will need to enable or disable cookies again.