TOS9300 – Firmware Updater

Firmware Updater

This software is to update the firmware version of the TOS9300 Series.

Ver.

1.17

Last update

2020/9/1

Configuration

Applied Products: TOS9300 Series – All models

Notes

How to use

See System Settings-Updating in the User’s Manual.

No Package Found!

    Download

    Release Note

    Firmware Version 1.19
    2023-1-16

    IFC1.08.0138/IOC1.09.0114/FPGA1.04.0007

    Resolved issues
    Fixed the problem that a different network is set when the power is restarted after performing a specific operation in the meter mode (Meter) of the leakage current test (LC).
    Firmware Version 1.18
    2022-9-06

    IFC1.08.0138/IOC1.08.0112/FPGA1.04.0007

    Improvement
    Improved the stability at long-term operation.
    Resolved issues
    Fixed the character display.
    Firmware Version 1.17
    2020-8-19

    FPGA1.04.0007/IOC1.07.0105/IFC1.07.0133

    Improvement
    Improved the UI for Memory.
    Improvement
    Improved the peak accuracy in LC.
    Improvement
    Improved the over range processing of the Meter mode.
    Improvement
    Improved the behavior to reduce the operation load of FPGA.
    Firmware Version 1.15
    2020-5-19

    FPGA1.03.0006/IOC1.06.0100/IFC1.06.0131

    Resolved issues
    Fixed the problem that the Interlock and the Calibration were not working at the same time when it was set to ‘Protect’ at start-up.
    Resolved issues
    Fixed the character display.
    Resolved issues
    Fixed the problem that the operation is unstable when starting the ‘AUTO’ test by entering the command from the Memory screen at Local mode.
    Updated
    Changed not to allow the Test Level changing during the Faill Time.
    Improvement
    Improved the behavior at the Contack Check of EC.
    Improvement
    Improved the stability at long-term operation.
    Firmware Version 1.14
    2020-1-17

    FPGA1.02.0005/IOC1.05.0093/IFC1.05.0130

    Resolved issues
    Fixed the problem that the response of auto test step number for 'RES:REM?' query was always '1'.
    Resolved issues
    Fixed the problem that C-FAIL was saved as L-FAIL in Memory.
    Resolved issues
    Fixed the problem that Calibration Protection was not working at power-on.
    Resolved issues
    Fixed the problem that the previous test condition was recalled at one time when the auto test memory was recalled through the signal I/O interface.
    Improvement
    Improved the behavior when 'IP Address Method' is set to 'Static'.
    Improvement
    Improved the firmware updater to display its version on the panel.
    Improvement
    Increased the memory items saved in Memory.
    Improvement
    Displays 'CHECK' on the panel during checking the scanner contact (Contact Check).
    Updated
    Changed the panel display from 'error' to 'over' when the measurement result exceeds the measurement range.
    Updated
    Changed the maximum period of date setting to 2037.
    Updated
    Changed not to allow the Test Level key pressing during the rise time.
    Resolved issues
    Fixed the problem that Calibration Protection was released by pressing STOP key even it was set to 'Enable'.
    Firmware Version 1.12
    2019-7-19

    IFC1.03.0123/IOC1.04.0077/FPGA1.02.0005

    Resolved issues
    Fixed the problem that was preventing tests from starting in rare cases when using the HP01A/02A-TOS.
    Resolved issues
    Fixed the problem that was preventing the next step from being performed in rare cases in an AUTO test even when FAIL Judgment was set to continue.
    Resolved issues
    Fixed the problem that was causing the measurement screen to display an error when an AUTO test was performed after a single test.
    Resolved issues
    Fixed the problem that was preventing an ACW test from being performed in rare cases even when a test was started by specifying an IR or ACW test.
    Improvement
    Improved the flickering measurement data display.
    Resolved issues
    Fixed the problem that was causing the instrument to return to READY immediately after a U-Fail occurrence in an AUTO test.
    Resolved issues
    Fixed the problem that was preventing the protection function from being activated even in an event of an overload during rising.
    Resolved issues
    Fixed the problem that was causing Over Heat protection and Power Supply protection to be misjudged in rare cases when long-term operation was performed.
    Resolved issues
    Fixed the problem that was causing the STOP bit to be set in the OPER:TEST register even when STOP was not pressed.
    Resolved issues
    Fixed the problem that was preventing the voltage rise rate judgment function from operating normally in rare cases.
    Resolved issues
    Fixed the problem that was causing L-FAIL to occur in rare cases when an IR test was started with the output open.
    Resolved issues
    Fixed the problem that was causing the instrument to operate at 1/10 of the specified upper and lower judgment values in an IR test during an auto test.
    Resolved issues
    Fixed the problem that was preventing tests from being performed according to the steps when a test was re-executed after releasing a protection function that was activated during an auto test.
    Improvement
    Improved the internal communication reliability.
    Firmware Version 1.10
    2019-5-08

    IFC1.07.0107

    Improvement
    Changed the digits and units displayed during measurement * Only the display was changed (the accuracy was not changed).
    Resolved issues
    Changed the digits and units displayed during measurement * Only the display was changed (the accuracy was not changed).
    Resolved issues
    The test time was set to 1 second. Fixed the problem that was causing the test time to be set to 0.2 seconds when the power was restarted.
    Resolved issues
    Fixed the problem that was preventing auto tests from completing when pass judgments were made (the problem kept repeating the test).
    Resolved issues
    Fixed the problem that was causing unused bits to be returned in response to the STAT:OPER:TEST:COND? command.
    Improvement
    Fixed the problem that was causing errors with the FUNC PROG command.
    Firmware Version 1.01
    2019-1-29

    IFC1.00.0092

    Resolved issues
    Fixed the problem that was preventing the instrument from starting during discharge.

    Please feel free to contact us with any questions or concerns.

    Download a Product Catalog View List
    Request a Quotation Contact Us
    close

    Search

    preloader