Fixed the issue where an interrupt operation done while saving a results file causes an error after testing is completed in HA mode.
Ver2.7.4
2024-8-29
HarmoCapture 3
Resolved issues
Fixed the issue where the test impedance display became inconsistent with that of KHA3000 when the wiring method was changed for IEC61000-3-11.
Resolved issues
Fixed the issue where 0 was not displayed on the SD006-KHA side when 0 was input as the specified value of test impedance from KHA3000 for IEC61000-3-11.
Resolved issues
Fixed the issue where multiple standards were selected when canceling the switching of reference impedance when selecting other standards from JIS61000-3-2 2019.
Resolved issues
Fixed the issue where the message “Reference impedance is not powered. Switch to manual?” was continuously displayed when SD006-KHA was started for IEC61000-3-12 when using line impedance.
Resolved issues
The system was fixed to display the standard value of the reference impedance for the wiring method at the start of the test for JIS61000-3-2.
Resolved issues
Fixed the issue where the impedance value was unable to be set to “Un-use (Bypass)” when set to remote with the reference impedance “Z4” or “Z5” selected for IEC61000-3-3.
Resolved issues
Fixed the issue where the control setting of reference impedance might not be applied correctly when a test condition file is opened in Vf mode.
Ver2.7.3
2024-6-20
HarmoCapture 3
Resolved issues
Fixed the issue where a test could not be started if, in JIS 61000-3-2, the ref impedance control was set to "Remote", and the impedance value to "Un-use (Bypass)".
Ver2.7.2
2024-3-25
HarmoCapture 3
Resolved issues
Fixed the issue where an error occurs when setting the voltage of PCR-LE or PCR-LE2 via KHA.
Resolved issues
Fixed the issue where an error occurs when changing the limitation standard in specific procedures.
2024-3-25
HA File Analyzer 3
Resolved issues
Fixed the issue where the repeatability confirmation page and report print display differed.
Ver2.7.1
2023-6-12
HarmoCapture 3
Resolved issues
Fixed the issue where the voltage range of channel L1 was not switched to a fixed range when a channel link was set independently.
Improvement
Wiring settings for the AC power supply were added to the AC power supply control pane.
2023-6-12
HA File Analyzer 3
Resolved issues
Fixed the issue where the item name was displayed as “Rated Voltage (Ui)” in the setting list regardless of the equipment classification when outputting IEC 61000-3-12 test reports.
2023-6-12
Vf File Analyzer 3
Resolved issues
Fixed the issue where the current range in the test report displays L1 range for all channels.
Resolved issues
Fixed the issue where the PLL source was displayed as “EXT L1” in the test report when loading the test data with the PLL source set to EXT.
The limitation standard was changed as follows from KHA3000 Ver3.50. IEC61000-3-2 Ed4.0 to IEC61000-3-2 Ed5.0 / JIS C 61000-3-2 2011 to JIS C 61000-3-2 2019 / IEC61000-3-3 Ed3.0 to IEC61000-3-3 Ed3.1 / IEC61000-3-11 Ed1.0 to IEC61000-3-11 Ed2.0 SD006-KHA was also changed according to it.
New Feature
Vf : When the measurement technical standard was IEC61000-4-15Ed2.0, the judgment limit value option display was changed into "Tmax" from "d(t) >3.3%."
New Feature
HA : When IEC61000-3-12 or JISC61000-3-2 were chosen and the start of test was carried out, next message came to come out. Test cannot start because the "Check of AC power supply" at xx has not been performed yet. If "Check of AC power supply" is performed once, this message will not come out.
Ver2.6.0
2021-4-08
HarmoCapture 3
New Feature
This version operates also in AC power supply PCR-WEA series. (Only Direct connection)
Resolved issues
There was a problem of which voltage range of the application was sometimes not the same as the voltage range of AC power supply.
Resolved issues
If output is turned on by AC power supply control pane, the group box of voltage range will gray out. However, when Limitation Standard etc. were changed, there was a problem of which gray out was canceled.
Resolved issues
Zref was chosen as test impedance by Vf mode IEC61000-3-11. However, there was a problem as which Z5 is displayed in the check dialog at the time of test start.
Resolved issues
When report printing of the test result file in Vf mode was carried out, there was a problem whose error occured and broke pdf.
2021-4-08
HA File Analyzer 3
Resolved issues
When printing report, there was a problem of which numerical value was printed instead of "Reference Impedance ---"
Ver2.5.4
2020-12-22
HarmoCapture 3 Vf mode
Resolved issues
When the number of times of d measurement was made into 24 times, there was a problem on which the report pdf of the test result broke. Other number of times was OK.
Resolved issues
The drop down list of the number of times of d measurement has 1. However, it cannot be chosen. 1 was removed from the drop down list. According to the operation manual, it is 3 to 24 times.
Resolved issues
In a setup of the nominal voltage, there is radio button of the nominal voltage 230V and the specified value. In the change of these radio buttons, there was a problem on which the nominal voltage command is not sent to KHA3000.
Ver2.5.3
2020-4-06
HarmoCapture 3
Resolved issues
There was a problem in which the error occurred at the time of footer creation of pdf, and incomplete pdf was made.
Ver2.5.2
2019-1-07
HarmoCapture 3
Resolved issues
Although check of the radio button of Z3 and Z5 was correct when the test condition file is read by Vf, there was a problem to which impedance value was not changed.
Resolved issues
After operation was carried out by "AC Power Supply - Use", when AC Power Supply was not connected, there was a problem to which an error occurred on launch of this app.
Updated
Operability of time setting of "minute + second is not 0" was corrected. If time is 0, message will be displayed when test start.
Ver2.5.1
2018-8-29
HarmoCapture 3
New Feature
AC power supply PCR-WE series have been added to the operation.
Ver2.5.0
2017-5-26
HarmoCapture 3
New Feature
For HA limitation standards, "IEC61000-3-2 Ed4.0" have been added.
New Feature
For Flicker limitation standards, "IEC61000-3-3 Ed3.0" have been added.
2017-5-26
HA File Analyzer 3
New Feature
For HA limitation standards, "IEC61000-3-2 Ed4.0" have been added.
2017-5-26
Vf File Analyzer 3
New Feature
For Flicker limitation standards, "IEC61000-3-3 Ed3.0" have been added.
Ver 2.4.2
2016-10-03
HarmoCapture 3
Resolved issues
There was a problem which does not conform to IEC61000-4-15 Ed2.0. The steady state and d measurement algorithm were changed.
Improvement
Contents and user's manual was updated.
2016-10-03
HA File Analyzer 3
Improvement
Contents and user's manual was updated.
2016-10-03
Vf File Analyzer 3
Improvement
Contents and user's manual was updated.
Ver 2.4.1
2016-1-18
HarmoCapture 3
Resolved issues
There was a problem in which an error occurs, when PCR-LE(2) operated by 1P2W, it sent the display command of 3 phases.
Resolved issues
When a test was started, there was a problem in which voltage range and current range of KHA3000 changes to the maximum. This occurs only in Ver2.4.0.
2016-1-18
Vf File Analyzer 3
Resolved issues
There was a problem which has an error in the value of "Reference Impedance" using the comment of file.
Ver 2.4.0
2015-11-17
HarmoCapture 3
New Feature
Connection of PCR series was only via KHA3000. In addition to it, direct connection from a PC was added.
New Feature
The added button deletes all records of Combo boxes of the Report setting form.
2015-11-17
HA File Analyzer 3
New Feature
The added button deletes all records of Combo boxes of the Report setting form.
2015-11-17
Vf File Analyzer 3
Resolved issues
When the small voltage acquired by L2 & L3 of 1P2W and L3 of 1P3W was drawn on the graph of a report, there was a problem in which the scale of axis was not displayed.
Resolved issues
When the line-voltage 230V was tested by 1P3W and nominal voltage was set as 230V, there was a problem in which the percent value of voltage variation became about 1/1.73.
New Feature
The added button deletes all records of Combo boxes of the Report setting form.
Ver 2.3.7
2015-1-28
HarmoCapture 3
Resolved issues
There was a problem which does not remember the impedance value in HA/Vf change and standard change when the manual mode that does not use the LIN-JF series.
Ver 2.3.6
2014-8-20
HarmoCapture 3
Resolved issues
When Ch-Link was "independent" in 1P2W and the range of L2 and L3 was switched, there was a problem of a command error.
Resolved issues
There was a problem which has the same value with different model in the list of combo boxes of Rated Current (Iequ), Ref Fund Current (I1) and Ref Current (Iref) of IEC61000-3-12.
Resolved issues
There was a problem in which change of meas time is not reflected.
2014-8-20
HA File Analyzer 3
Resolved issues
There was a problem which has mistake in the value of Reference Current Iref (ave) in text conversion.
Improvement
Setting like 1-40 was permitted at a setup of the order of the harmonics trend of a text conversion option.
Ver 2.3.5
2017-7-18
HarmoCapture 3 / HA File Analyzer 3
Resolved issues
Each phase page of report was as follows in 3P3W. ・Real Power (max) --- ・Apparent Power (max) --- ・Reactive Power (max) --- ・Power Factor (max) --- ・Sequ --- ・Ssc ---
Resolved issues
There was a problem that a setup of impedance value changes when start of test by Vf.
Resolved issues
There was a problem that when save of text then error message is displayed in JIS C 61000-3-2 Class A "600W Air-conditioner:Yes".
Ver 2.3.4
2014-6-27
HarmoCapture 3
Resolved issues
There were mistakes in the following value of a report on IEC61000-3-12 (2011). ・Reference Current Iref (ave) ・Reading (In/Iref) Ave[%] ・Reading (In/Iref) Max[%] ・Judge ・THC/Iref[%] ・PWHC/Iref[%] ・Minimum Rsce
2014-6-27
HA File Analyzer 3
Improvement
The bar graph of primary current is not displayed in 2D / 3D harmonics of IEC61000-3-12.
Ver 2.3.2
2014-3-17
HarmoCapture 3
Resolved issues
There was a problem that an exception-error occurred by a find resource, when no instruments connected, .
Resolved issues
When harmonics-flicker mode was switched, there was a problem that impedance does not become a bypass.
Improvement
Corresponded to control of LIN(Z5) of the flicker standard more than 16A.
Ver 2.3.1
2013-12-20
HarmoCapture 3
Resolved issues
There was a problem of application starting error, when there was no registry information.
Ver 2.3.0
2013-12-06
HarmoCapture 3
New Feature
For HA limitation standards, "JIS C 60200-3-2 2011" have been added.
New Feature
For HA limitation standards, "IEC 60200-3-12 2011" have been added.
Improvement
Measurement time was extended to 9600 seconds.
New Feature
Corresponded to remote control of LIN-JF series.
Ver 2.2.2
2012-9-14
HarmoCapture 3
New Feature
AC power supply PCR-LE series have been added to the operation.
Ver 2.2.1
2011-5-16
HarmoCapture 3
Improvement
Contents and user's manual was updated.
2011-5-16
HA File Analyzer 3
Improvement
Contents and user's manual was updated.
2011-5-16
Vf File Analyzer 3
Improvement
Contents and user's manual was updated.
Ver 2.2.0
2010-12-20
HarmoCapture 3
Improvement
For HA limitation standards, "IEC61000-3-2 Ed3.0 A2" and measurement technic "IEC61000-4-7 Ed2.0 A1" have been added.
New Feature
Reference impedance is now selectable to "Use" or "Un-use" when JIS2005 selected.
Improvement
For Flicker limitation standards, "IEC61000-3-3 Ed2.0" and measurement technic "IEC61000-4-15 Ed2.0" have been added.
2010-12-20
HA File Analyzer 3
New Feature
Confirmation of AC power source have been added. This is applied to screen display and reports.
Improvement
Only when HA limitation standard selected IEC61000-3-2 Ed3.0 A2, calculation denominator of a repeatability is the limit value. Less than 5% is Pass.
Ver 2.1.4
2010-9-16
HarmoCapture 3
New Feature
For HA, IEC 61000-3-2 Ed3.0 A2 is added to limitation standard, and IEC 61000-4-7 Ed2.0 A1 is added to measurement standard.
New Feature
For HA, Only when Class C is selected under the limitation standard IEC 61000-3-2 Ed3.0A2, 3rd/5th/current-waveform are now displayed at applied limitation values.
2010-9-16
HA File Analyzer 3
New Feature
A description figure of current waveform requirement is now drawn at V/I waveform when a result file where the applied limit values are 3rd/5th/current-waveform in Class-C. This is applied to screen display and reports.
New Feature
The 2D harmonic current auto-scale is now selectable to "Meas Value" or "Meas and Limit". In "Meas and Limit", the appropriate scale is calculated from measurement values and limit values, therefore the printing problem that the limit values were too high to print has been fixed.
2010-9-16
Vf File Analyzer 3
New Feature
When report printing, the latest EN61000-3-3 (2008) / *IEC61000-3-3:Ed2.0 (2008) is now printed for "Type of test."
Ver 2.1.3
2010-1-08
HarmoCapture 3
Resolved issues
In spite of using old-formatted files (file version: HA result file 1.03 or prior, Vf result file 1.01 or prior) which do not contain local standard name, there was a problem that generates an error when selecting "Read Alias from this file" on the comment selection.
Resolved issues
There was a problem that the program fell into infinite loop, when printing with "Do not display PDF Over Write Message" option on report settings is being checked.
Ver 2.1.2
2009-10-15
HarmoCapture 3
Resolved issues
In the vf report of manual switching d-measurement, there was a problem that the report prints out "Pst Measurement Time" and "Pst Measurement Count", which should be "d Measurement Time" and "d Measurement Count".
Improvement
English contents and user's manual can be now referenced from the Help menu.
2009-10-15
HA File Analyzer 3
Improvement
English contents and user's manual can be now referenced from the Help menu.
2009-10-15
Vf File Analyzer 3
Resolved issues
There was a problem that graphs could not be drawn in the report when the test was 60 seconds or longer.
Improvement
English contents and user's manual can be now referenced from the Help menu.
Ver 2.1.1
2009-3-23
HarmoCapture 3
New Feature
On the modeless sub-form (where you can operate main-form without closing), current input terminal and scaling setting functions are added.
New Feature
DC offset setting function is added.
New Feature
The new function, which allows to select retry operation for each segment at IEC 61000-3-3 (Manual Sw), is added.
Improvement
On the main form, the new indicators, where scaling on/off state and current input terminal settings can be seen, have been added.
Improvement
Newly provided a choice, which allows to specify other folder and file name when a report PDF having the same file name already existing.
Improvement
As well other existing conditions, the warning dialog at test start operation also appears: »When the power is negative due to wrong connection, or »When the sensor could not be detected with the current input sensor enabled
Resolved issues
There was a problem that margin could not set at IEC 61000-3-12 and IEC 61000-3-11.
Resolved issues
There was a problem that 600W Air-conditioner setting might not be loaded at JIS 61000-3-2 Class A.
2009-3-23
HA File Analyzer 3
Improvement
Newly supported local standard names for printing, allowing to create reports with an arbitrary standard name such as GB standard.
Improvement
Newly provided a choice, which allows to specify other folder and file name when a report PDF having the same file name already existing.
Improvement
Test results are now displayed beneath the toolbar, allowing to view them without tab switching.
Improvement
Newly added Test Result tab, where the max and limit values of POHC, the minimum Rsce of IEC 61000-3-12, Sequ, Ssc, Z values can be displayed.
Improvement
Newly added the following drawing conditions at V/I Waveform of Report Setting. Under IEC 61000-3-12: »Minimum Rsce Maximum »THD Maximum frame »PWHD Maximum »Selected frame
Improvement
As result files acquired by KHA3000 Ver2.0 or later embed horizontal scale values of V/I waveform, the display now uses these embedded values. As for result files acquired earlier than KHA3000 Ver2.00, the horizontal scale values of V/I waveform are decided by calculation.
Improvement
As for limit values, display of calculated 0A and unapplied case (---) are now distinguished being displayed differently.
Resolved issues
There was a problem that the vector lines were displayed slightly longer than real values.
Resolved issues
There was a problem that an error was generated when opening a result file created under IEC 61000-3-12 single-phase test.
Resolved issues
There was a problem that an error was generated when opening a result file created by earlier than KHA3000 Ver2.00.
2009-3-23
Vf File Analyzer 3
Improvement
Newly supported local standard names for printing, allowing to create reports with an arbitrary standard name such as GB standard.
Improvement
Newly provided a choice, which allows to specify other folder and file name when a report PDF having the same file name already existing.
Resolved issues
There was a problem that Zmax value was not printed on IEC61000-3-11 reports.
Resolved issues
There was a problem that d-margin was missing on IEC61000-3-11 display and reports.
Ver 2.0.0
2008-11-07
HarmoCapture 3
Resolved issues
There was a problem that PCR-LA did not return to LOCAL when the application exits while PCR-LA OUTPUT is kept ON.
Resolved issues
There was a problem that Σ current measurement value was 1 digit smaller than real current value.
Resolved issues
There was a problem that the limit display of dmax, dc, d(t)>3.3% for each channel on IEC 61000-3-3 (Manual Sw) report output was shown as "-.--".
Resolved issues
There was a problem that 2D Harmonics on IEC 61000-3-12 report output did not show in red even if exceeding limit values.
2008-11-07
HA File Analyzer 3
New Feature
The new application is added.
2008-11-07
Vf File Analyzer 3
New Feature
The new application is added.
Ver 1.0.1
2008-4-11
HarmoCapture 3
New Feature
The first release
Please feel free to contact us with any questions or concerns.
Cookies will be used to track which website is being visited. On our website, we certify the information that can be accessed by visitors using cookies. We also track viewed information by users to understand your interest and provide you with a better content. You will be able to disable cookie itself, you may choose the way to use personal identification whenever you use service that requires authentication.
Strictly Necessary Cookies
Strictly Necessary Cookie should be enabled at all times so that we can save your preferences for cookie settings.
If you disable this cookie, we will not be able to save your preferences. This means that every time you visit this website you will need to enable or disable cookies again.
Cookies will be used to track which website is being visited. On our website, we certify the information that can be accessed by visitors using cookies. We also track viewed information by users to understand your interest and provide you with a better content. You will be able to disable cookie itself, you may choose the way to use personal identification whenever you use service that requires authentication.
Strictly Necessary Cookies
Strictly Necessary Cookie should be enabled at all times so that we can save your preferences for cookie settings.
If you disable this cookie, we will not be able to save your preferences. This means that every time you visit this website you will need to enable or disable cookies again.